Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Over-Head Ceiling Tube Support Recalled by SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION Due to Eight fixing bolts at the base of the...

Date: April 10, 2023
Company: SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION directly.

Affected Products

Over-Head Ceiling Tube Support, CH-200, CH-200M, Used with RadSpeed and RadSpeed PRO, Sonial Vision safire, Versa and G4 units, FluoroSpeed 300 and X1 units

Quantity: 1836

Why Was This Recalled?

Eight fixing bolts at the base of the X-ray ceiling tube column assembly may be loose or missing so service is required to install a cable, bracket, and label behind the bracket, indicating when service is required, to eliminate the risk of the device falling.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION

SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION has 6 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report