Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

uEXPLORER PET/CT Diagnostic Imaging System - Product Usage: This system Recalled by Shanghai United Imaging Healthcare Co., Ltd. Due to 1) 1) Potential intermittent issue may cause image...

Date: July 16, 2020
Company: Shanghai United Imaging Healthcare Co., Ltd.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Shanghai United Imaging Healthcare Co., Ltd. directly.

Affected Products

uEXPLORER PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

Quantity: 1 unit distributed to Texas

Why Was This Recalled?

1) 1) Potential intermittent issue may cause image data problems, may lead to artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external interference; 3) Metal edge overlap of mylar strip in gantry may create sharp edge. May lead to repeat scanning and abrasions.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Shanghai United Imaging Healthcare Co., Ltd.

Shanghai United Imaging Healthcare Co., Ltd. has 21 total recalls tracked by RecallDetector.