Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Computed Tomography X-Ray System Recalled by Shanghai United Imaging Healthcare Co., Ltd. Due to A potential issue of an incorrect CT scan...

Date: August 5, 2020
Company: Shanghai United Imaging Healthcare Co., Ltd.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Shanghai United Imaging Healthcare Co., Ltd. directly.

Affected Products

Computed Tomography X-Ray System, Model uCT 760, Rx, CE, GTIN: 06971576831043 - Product Usage: ntended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

Quantity: 5

Why Was This Recalled?

A potential issue of an incorrect CT scan delay timer to be displayed on the gantry Digital Display Panel (DDP) during multi-phase contrast scanning. This could lead the operator to receive unnecessary exposure to radiation.

Where Was This Sold?

This product was distributed nationwide across all 50 states.