Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

IRON assay Recalled by Sentinel CH SpA Due to A high recovery of Iron assay has been...

Date: September 29, 2023
Company: Sentinel CH SpA
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Sentinel CH SpA directly.

Affected Products

IRON assay, Reference Numbers 6K95-41 and 6K95-30

Quantity: 4394 kits

Why Was This Recalled?

A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for Iron Products (Alinity c Iron Reagent REF. 08P3920) to be used on Abbott Alinity Instrument and to 30% for Iron Products (IRON REF. 6K95-41 and IRON REF. 6K95- 30) to be used on Abbott Architect Instrument. When using the affected lots, there is a potential for delay in reporting patient results due to QC failure, also a potential for falsely elevated results.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Sentinel CH SpA

Sentinel CH SpA has 7 total recalls tracked by RecallDetector.