Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Alinity c Lithium Reagent Kit- intended for the quantitation of Recalled by Sentinel CH SpA Due to Added an additional SmartWash parameter for the Lithium...

Date: June 28, 2019
Company: Sentinel CH SpA
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Sentinel CH SpA directly.

Affected Products

Alinity c Lithium Reagent Kit- intended for the quantitation of lithium in serum or plasma using respectively the ARCHITECT c Systems and the Alinity c analyzer List Number: 08P5320

Quantity: 1217 kits

Why Was This Recalled?

Added an additional SmartWash parameter for the Lithium and Alinity c Lithium Reagent Kit, catalogue number 8L25-30 and 08P5320 to prevent the potential for carryover from the ARCHITECT Lactate Dehydrogenase (LDH) reagent (LN 2P56) may result infalsely elevated lithium patient results

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Sentinel CH SpA

Sentinel CH SpA has 7 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report