Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MobileDiagnost wDR Recalled by SEDECAL SA Due to While cleaning or disinfecting the system, if the...

Date: October 14, 2021
Company: SEDECAL SA
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact SEDECAL SA directly.

Affected Products

MobileDiagnost wDR, Model numbers 712001, 712002, 712004, 712006, and 712007

Quantity: 851 devices

Why Was This Recalled?

While cleaning or disinfecting the system, if the process is not followed as described in the manual and excessive liquid is used or sprayed directly on the system, it could cause a short circuit of the fine positioning control button. This would result in the system moving on its own in a very slow speed forward circular motion.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About SEDECAL SA

SEDECAL SA has 53 total recalls tracked by RecallDetector.