Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Arial mobile application Recalled by Securitas Healthcare LLC Due to The Arial Mobile Application utilizing Apple iOS 17.4...

Date: April 11, 2024
Company: Securitas Healthcare LLC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Securitas Healthcare LLC directly.

Affected Products

Arial mobile application, model numbers 54640, 54630, and 54630-SMA, that utilizes Apple iOS 17.4 and Apple iOS 17.5, a component of the Arial Emergency and Nurse Call System.

Quantity: 661 sites with mobile application user licenses

Why Was This Recalled?

The Arial Mobile Application utilizing Apple iOS 17.4 and iOS 17.5 is frequently restarting causing the app to receive alerts for all pending alarm notifications and other events when the app restarts, experience difficulty in placing and maintaining staff-to-staff phone calls, and frequently appears offline while the app restarts.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Securitas Healthcare LLC

Securitas Healthcare LLC has 10 total recalls tracked by RecallDetector.

Related Recalls