Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Curaplex Field Cric Kit with ET Tube Recalled by Sarnova HC, Llc Due to It was determined that the Endotracheal Tube Holder...

Date: November 6, 2023
Company: Sarnova HC, Llc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Sarnova HC, Llc directly.

Affected Products

Curaplex Field Cric Kit with ET Tube, Item Number 36611. The kit provides healthcare professionals with an assortment of supplies to perform a cricothyrotomy.

Quantity: 1063 kits

Why Was This Recalled?

It was determined that the Endotracheal Tube Holder included in certain kits is not compatible with a cricothyrotomy procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Sarnova HC, Llc

Sarnova HC, Llc has 4 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report