Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Curaplex Cricothyrotomy Field Kit with Instruments Recalled by Sarnova HC, Llc Due to It was determined that the Endotracheal Tube Holder...

Date: November 6, 2023
Company: Sarnova HC, Llc
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Sarnova HC, Llc directly.

Affected Products

Curaplex Cricothyrotomy Field Kit with Instruments, Kwik Cric MVP, 6.0mm, Item Number 351631i-A. The kit provides healthcare professionals with an assortment of supplies to perform a cricothyrotomy.

Quantity: 15 kits

Why Was This Recalled?

It was determined that the Endotracheal Tube Holder included in certain kits is not compatible with a cricothyrotomy procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Sarnova HC, Llc

Sarnova HC, Llc has 4 total recalls tracked by RecallDetector.

Related Recalls