Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

uPath software 1.1 and uPath software 1.1 IVD-INSW Recalled by Roche Molecular Systems, Inc. Due to When a user creates a measurement tool annotation...

Date: November 23, 2020
Company: Roche Molecular Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Roche Molecular Systems, Inc. directly.

Affected Products

uPath software 1.1 and uPath software 1.1 IVD-INSW, Material numbers 08991871001 and 09077332001

Quantity: 27

Why Was This Recalled?

When a user creates a measurement tool annotation in the uPath Enterprise software version 1.1, the measurement value is incorrectly calculated when the measurement is viewed in "Split View" viewing mode and the slides within the case are scanned at different magnifications.

Where Was This Sold?

Domestic distribution to Maryland and Michigan. International distribution to France, Greece, Italy, Poland, Turkey, Brazil, India, and Singapore.