Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

COBAS AmpliPrep Sample Tube Input Barcode Clips Recalled by Roche Molecular Systems, Inc. Due to Certain lots of COBAS AmpliPrep (CAP) S-tube Input...

Date: February 1, 2012
Company: Roche Molecular Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Roche Molecular Systems, Inc. directly.

Affected Products

COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. Input S-tube barcode clips hold sample input tubes in the COBAS AmpliPrep Instrument (CAP) sample racks and have a barcode to identify the sample.

Quantity: 2,132 boxes

Why Was This Recalled?

Certain lots of COBAS AmpliPrep (CAP) S-tube Input (Box of 12 bags 12x24 S-tube and 12x24 S-clip) can cause the sample input tubes to stick in the barcode clips. As a consequence, the CAP instrument will lift both the sample input tube together with the barcode clip from the sample rack, resulting in an aborted run. No erroneous results would be produced if the issue occurs, but samples in proce

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected