Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

cobas 6800/8800 system Hepatitis Viral B DNA Detection Product Usage: Recalled by Roche Diagnostics Operations, Inc. Due to Roche has received four reports of injury caused...

Date: July 15, 2016
Company: Roche Diagnostics Operations, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Roche Diagnostics Operations, Inc. directly.

Affected Products

cobas 6800/8800 system Hepatitis Viral B DNA Detection Product Usage: The cobas p512 pre-analytical system is a computer controlled fully automatic system for sorting of open and closed barcoded and centrifuged sample tubes. It includes modules for registration and decapping of sample tubes, liquid level detection (optional) and sample quality detection (optional), as well as recapping of sample tubes (optional).

Quantity: 9014 in total

Why Was This Recalled?

Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Roche Diagnostics Operations, Inc.

Roche Diagnostics Operations, Inc. has 127 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report