Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used Recalled by Roche Diagnostics Operations, Inc. Due to Roche investigation into complaints found Accu-Chek Inform II...

Date: March 6, 2013
Company: Roche Diagnostics Operations, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Roche Diagnostics Operations, Inc. directly.

Affected Products

CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to quantitatively determines prothrombin time ("PT").

Quantity: 4467 both devices

Why Was This Recalled?

Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards. Errors have been confirmed with the EAN 13 barcodes which should not be used for patient or operator identification.

Where Was This Sold?

This product was distributed nationwide across all 50 states.