Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Elecsys Vitamin D total II Recalled by Roche Diagnostics Corporation Due to The device may give a falsely elevated result...

Date: April 27, 2018
Company: Roche Diagnostics Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Roche Diagnostics Corporation directly.

Affected Products

Elecsys Vitamin D total II, Vitamin D Test System, Material number 07028148190 Product Usage: This assay is intended for the quantitative determination of total 25-hydroxyvitamin D in human serum and plasma. This assay is to be used as an aid in the assessment of vitamin D sufficiency in adults. The electrochemiluminescence binding assay is intended for use on the immunoassay analyzer.

Quantity: 347 units

Why Was This Recalled?

The device may give a falsely elevated result that is non-reproducible. If vitamin D level is monitored by laboratory tests and adapted based on the result, a dosage reduction of vitamin D supplementation cannot be excluded, and as a consequence, vitamin D deficiency may develop over time.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Roche Diagnostics Corporation

Roche Diagnostics Corporation has 39 total recalls tracked by RecallDetector.

Related Recalls