Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

cobas c 111 with ISE analyzers analyzer used in the Recalled by Roche Diagnostics Corporation Due to COBAS INTEGRA¿ 400 plus and cobas c 111...

Date: December 14, 2018
Company: Roche Diagnostics Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Roche Diagnostics Corporation directly.

Affected Products

cobas c 111 with ISE analyzers analyzer used in the following assays: a)Creatinine plus ver.2 Catalog Number: 05401470190; b)Triglycerides Catalog Number: 04657594190 - Product Usage: Tina-quant Apolipoprotein A-1 ver.2 A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatment of lipid disorders and atherosclerosis Tina-quant Albumin Gen.2 Immunoturbidimetric assay for the quantitative, in vitro determination of albumin in human urine, serum, plasma, and CSF on Roche/Hitachi cobas c systems

Quantity: 137 units

Why Was This Recalled?

COBAS INTEGRA¿ 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Roche Diagnostics Corporation

Roche Diagnostics Corporation has 39 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report