Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Revive Reusable Bladder Support Recalled by Rinovum Women's Health Due to The product is unable to meet the use...

Date: May 2, 2019
Company: Rinovum Women's Health
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Rinovum Women's Health directly.

Affected Products

Revive Reusable Bladder Support, Model No. 0025 - Product Usage: The Revive bladder support is intended for temporary management of urinary leakage caused by stress urinary incontinence (SUI) in women, 18 years and older. The device comes in a sealed box, and contains a reusable bladder support and applicator, and disposable strings and travel case. The device is nonperishable and does not have an expiration date.

Quantity: 1024 (180 distributed from distributor)

Why Was This Recalled?

The product is unable to meet the use life of 31 cycles and may split or crack at the base of the "petals" after repeated use but within the use life of the product (31 uses).

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Rinovum Women's Health

Rinovum Women's Health has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report