Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Reuseable rotary blade/abrader Recalled by Richard Wolf Medical Instruments Corp. Due to Weak weld in the reusable rotary blade/abrader (burr)...

Date: November 25, 2015
Company: Richard Wolf Medical Instruments Corp.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Richard Wolf Medical Instruments Corp. directly.

Affected Products

Reuseable rotary blade/abrader, product number 899751504. The oval burr laterally hooded is an attachment or accessory to the air-powered device intended for use during surgical procedures to cut hard tissue or bone and soft tissue. General hospital and plastic surgery use.

Quantity: four units

Why Was This Recalled?

Weak weld in the reusable rotary blade/abrader (burr) may result in inner part breaks at the distal or proximal end making the product dysfunctional.

Where Was This Sold?

Distributed in Washington.

About Richard Wolf Medical Instruments Corp.

Richard Wolf Medical Instruments Corp. has 9 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report