Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Philips V60 Ventilators with Power Management PCBA part number 1055906 Recalled by Respironics California, LLC Due to Solder connection failure on Power Management printed circuit...

Date: March 5, 2020
Company: Respironics California, LLC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Respironics California, LLC directly.

Affected Products

Philips V60 Ventilators with Power Management PCBA part number 1055906

Quantity: 23581

Why Was This Recalled?

Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Respironics California, LLC

Respironics California, LLC has 46 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report