Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Philips V60 Ventilators Recalled by Respironics California Inc Due to Over time, low-frequency vibrations can cause the pins...

Date: April 24, 2017
Company: Respironics California Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Respironics California Inc directly.

Affected Products

Philips V60 Ventilators, Material No. 850008, 1053613, 1053614, 1053615, 1053616, 1053617, 1076715, 1076716, 1076717, DU1053617, U1053617. Old Material No. 850008, 989805628251, 989805612101, 989805613391, 989805613661, 989805611761, 989805627411, 989805627431, 989805627441, 989805616411, 989805636631.

Quantity: 30,516 units

Why Was This Recalled?

Over time, low-frequency vibrations can cause the pins within the female connectors on the Motor Controller to Data Acquisition Board ribbon cable to become partially displaced causing momentary high resistance that prevents or restricts data travelling through the cable during the high resistance event.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Respironics California Inc

Respironics California Inc has 7 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report