Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ReShape Integrated Dual Balloon System Recalled by Reshape Medical Inc Due to Due to a suspected misalignment of the connection...

Date: May 4, 2016
Company: Reshape Medical Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Reshape Medical Inc directly.

Affected Products

ReShape Integrated Dual Balloon System, Gen 1, Catalog No. RSM110, Model No. 01-0013-001 The ReShape Integrated Dual Balloon System is indicated for weight reduction when used in conjunction with diet and exercise, in obese patients with a Body Mass Index (BMI) of 30 40 kg/m2 and one or more obesity related comorbid conditions. It is indicated for use in adult patients who have failed weight reduction with diet and exercise alone.

Quantity: 56 units

Why Was This Recalled?

Due to a suspected misalignment of the connection between the proximal balloon cap of the implant and the delivery catheter. The possible misalignment is the result of an assembly error during manufacturing and may lead to leakage of saline solution during the balloon inflation procedure.

Where Was This Sold?

This product was distributed to 3 states: CA, NY, OH

Affected (3 states)Not affected

About Reshape Medical Inc

Reshape Medical Inc has 2 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report