Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Integra 3 Recalled by RESEARCH INSTRUMENTS LTD - SITE B Due to The units have been fitted with a potentially...

Date: March 11, 2016
Company: RESEARCH INSTRUMENTS LTD - SITE B
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact RESEARCH INSTRUMENTS LTD - SITE B directly.

Affected Products

Integra 3; Model Numbers: 6-54-100, 6-54-110, 6-54-150, 6-54-160 Product Usage: The Rl TDU500 Micromanipulator for Assisted Reproduction is used to accurately position microtools under a microscope for the techniques of Assisted Reproduction. The Rl Integra is a configuration comprising two TDU500 micromanipulators, two air syringes and temperature control system within a single housing.

Quantity: 380 units

Why Was This Recalled?

The units have been fitted with a potentially faulty power supply cable.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About RESEARCH INSTRUMENTS LTD - SITE B

RESEARCH INSTRUMENTS LTD - SITE B has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report