Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Refractor Suspension Arm (P/N: 525004 Recalled by Reliance Medical Products Inc Due to Product has been found with a missing Washer,...

Date: December 6, 2017
Company: Reliance Medical Products Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Reliance Medical Products Inc directly.

Affected Products

Refractor Suspension Arm (P/N: 525004, 05 and 525006, 07) The Intended use of this medical device, as part of system to store and position ophthalmological examination and diagnostic equipment, is to store and position for examination a device commonly known as a refractor.

Quantity: 933 ( of which 97 has been quarantined)

Why Was This Recalled?

Product has been found with a missing Washer, Retaining Ring, and/or Stop Screw. Recalling firm was notified by distributor that Refractor Suspension Arm fell on a patient s lower torso along with the arm housing

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Reliance Medical Products Inc

Reliance Medical Products Inc has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report