Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

SharePlan 1.1 Recalled by RAYSEARCH LABORATORIES AB Due to This notice concerns behavior of the SharePlan that...

Date: October 10, 2011
Company: RAYSEARCH LABORATORIES AB
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact RAYSEARCH LABORATORIES AB directly.

Affected Products

SharePlan 1.1, aka t-RayAutoplan version 1.2.1 (build number 1.3.1.10). Manufacturing name RayAutoplan, aka t-RayAutoplan, commercial name (TomoTherapy) SharePlan Intended use: t-RayAutoplan is a software that, based on the planned dose distribution from a TomoPlan treatment planning system, generates a family of individually optimized IMRT plans and presents their characteristics to the user in a GUI. Among the generated plans, the user selects and clinically approves a treatment plan and exports it electronically to DICOM, for subsequent treatment of the patient. The intended users of t-RayAutoplan shall be clinically qualified staff, such as medical physicists, medical doctors or dosimetrists.

Quantity: 20 units

Why Was This Recalled?

This notice concerns behavior of the SharePlan that may be unexpected and not clearly described in the user manual. This behavior has not caused patient mistreatment or other incidents. However, the described user actions must be adhered to for best agreement between treatment plan evaluation and the clinical outcome of the delivered treatment. The behavior appears, when importing RFA-files (.asc)

Where Was This Sold?

This product was distributed to 8 states: CA, KY, MI, NY, OK, OR, SD, WI

Affected (8 states)Not affected

About RAYSEARCH LABORATORIES AB

RAYSEARCH LABORATORIES AB has 113 total recalls tracked by RecallDetector.

Related Recalls