Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

RayStation/RayPlan 8B Service Pack 2 Recalled by RAYSEARCH LABORATORIES AB Due to There is a potential for erroneous propagation of...

Date: February 20, 2020
Company: RAYSEARCH LABORATORIES AB
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact RAYSEARCH LABORATORIES AB directly.

Affected Products

RayStation/RayPlan 8B Service Pack 2, Build Number 8.1.2.5, stand-alone software treatment planning system

Quantity: 1 system in the U.S.

Why Was This Recalled?

There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.

Where Was This Sold?

This product was distributed to 1 state: TN

Affected (1 state)Not affected

About RAYSEARCH LABORATORIES AB

RAYSEARCH LABORATORIES AB has 113 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report