Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

RayStation 4 Recalled by RAYSEARCH LABORATORIES AB Due to Two issues were found with the treatment plan...

Date: November 11, 2020
Company: RAYSEARCH LABORATORIES AB
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact RAYSEARCH LABORATORIES AB directly.

Affected Products

RayStation 4, 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, RayPlan 1, 2, 7, 8A, 8B, 9A, 9B, including all service packs. RayStation 10A and RayPlan 10A are also affected, but not the 10A service pack. (Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0, 9.1, 9.2, 10.0)

Quantity: 124

Why Was This Recalled?

Two issues were found with the treatment plan report: 1) There may be incorrect information for the percent of a region of interest with a clinical goal that is outside the dose grid; and 2) For RayStation and RayPlan 9B and 10A, 6D couch angles may be incorrect.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About RAYSEARCH LABORATORIES AB

RAYSEARCH LABORATORIES AB has 113 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report