Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Radiation Therapy Treatment Planning System Recalled by RAYSEARCH LABORATORIES AB Due to For a treatment plan consisting of multiple beam...

Date: May 30, 2016
Company: RAYSEARCH LABORATORIES AB
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact RAYSEARCH LABORATORIES AB directly.

Affected Products

Radiation Therapy Treatment Planning System, Model 5.0

Quantity: 491 units (187 domestically & 304 internationally)

Why Was This Recalled?

For a treatment plan consisting of multiple beam sets, the table for ROI plan dose statistics in the report may show the statistics for a beam set dose. The error can only be triggered when using a report template where statistics for a beam set dose is included as the last dose statistics prior to the plan dose statistics.