Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Brand Name: RayStation Intended as a Radiation Therapy Treatment Planning Recalled by RAYSEARCH LABORATORIES AB Due to Inconsistency in the use of Density uncertainty in...

Date: January 16, 2025
Company: RAYSEARCH LABORATORIES AB
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact RAYSEARCH LABORATORIES AB directly.

Affected Products

Brand Name: RayStation Intended as a Radiation Therapy Treatment Planning System. Model/Catalog Number: 7.0.0.19, 8.0.0.61, 8.0.1.10 Software Version: RayStation 7, 8A and 8A Service Pack 1

Quantity: 0

Why Was This Recalled?

Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About RAYSEARCH LABORATORIES AB

RAYSEARCH LABORATORIES AB has 113 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report