Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Randox Lipase (Colorimetric) Product Usage: A lipase test system is Recalled by Randox Laboratories Ltd. Due to According to the firm carry over was observed...

Date: August 2, 2016
Company: Randox Laboratories Ltd.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Randox Laboratories Ltd. directly.

Affected Products

Randox Lipase (Colorimetric) Product Usage: A lipase test system is a device intended to measure the activity of the enzyme lipase in serum and plasma. Lipase measurements are used in the diagnosis and treatment of diseases of the pancreas such as acute pancreatitis and obstruction of the pancreatic duct.

Quantity: 18

Why Was This Recalled?

According to the firm carry over was observed when the Lipase test is run directly before or after the LDL and Triglycerides assays. Customers are instructed to amend the running order for all instruments using these reagents to ensure Lipase is not place directly before or after LDL or Triglycerides. A correction was made to the IFU for the Lipase assay and all customers who purchased the Lipase kit in the USA were contacted with the updated instruction.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Randox Laboratories Ltd.

Randox Laboratories Ltd. has 72 total recalls tracked by RecallDetector.