Class III — Low Risk

Low risk — use of or exposure to this product is not likely to cause adverse health consequences.

Liquid Clinical Chemistry Control Recalled by Randox Laboratories Ltd. Due to A transcription error has occurred for the Randox...

Date: December 13, 2019
Company: Randox Laboratories Ltd.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Randox Laboratories Ltd. directly.

Affected Products

Liquid Clinical Chemistry Control, Catalogue Number: LAE4217, Size: 12 x 5 ml

Quantity: 2

Why Was This Recalled?

A transcription error has occurred for the Randox Lipase colorimetric value in Liquid Clinical Chemistry Control lot 1021 UE. The target and range are listed as 50 U/L (40 - 60) and should be 72U/L (58 - 86).

Where Was This Sold?

This product was distributed to 2 states: DE, NY

Affected (2 states)Not affected

About Randox Laboratories Ltd.

Randox Laboratories Ltd. has 72 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report