Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Liquid Assayed Specific Protein Control Level 3- In vitro diagnostic use Recalled by Randox Laboratories Ltd. Due to Concentration of Rheumatoid Factor has decreased in Specific...

Date: November 22, 2022
Company: Randox Laboratories Ltd.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Randox Laboratories Ltd. directly.

Affected Products

Liquid Assayed Specific Protein Control Level 3- In vitro diagnostic use, in the quality control of serum on clinical chemistry and immunoassay systems Catalog Number: PS2684

Quantity: 49 units US

Why Was This Recalled?

Concentration of Rheumatoid Factor has decreased in Specific Protein Control Level 1-3 causing delay in reporting results due to Quality Controls running low outside range

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Randox Laboratories Ltd.

Randox Laboratories Ltd. has 72 total recalls tracked by RecallDetector.

Related Recalls