Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Calibration Serum Level 3 - IVD For use as a Recalled by Randox Laboratories Ltd. Due to Target for Alkaline Phosphatase (ALP) for the AMP...

Date: November 20, 2023
Company: Randox Laboratories Ltd.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Randox Laboratories Ltd. directly.

Affected Products

Calibration Serum Level 3 - IVD For use as a Calibrator in clinical chemistry assays Catalog Number: CAL2351

Quantity: 108 kits

Why Was This Recalled?

Target for Alkaline Phosphatase (ALP) for the AMP optimized to IFCC 37¿C method has been mis-assigned for the RX Series instruments by approximately 10% on Quality Control and patient samples

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Randox Laboratories Ltd.

Randox Laboratories Ltd. has 72 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report