Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Calibration Serum Level 3 CAL2351 Recalled by Randox Laboratories Ltd. Due to CK Total in Calibration Serum Level 3, CAL2351,...

Date: November 20, 2023
Company: Randox Laboratories Ltd.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Randox Laboratories Ltd. directly.

Affected Products

Calibration Serum Level 3 CAL2351

Quantity: 269 kits

Why Was This Recalled?

CK Total in Calibration Serum Level 3, CAL2351, is running with a positive bias on RX Series instruments compared to other methods experience a shift in Quality Control and patient sample recovery of up to 13%.

Where Was This Sold?

This product was distributed nationwide across all 50 states.