Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Calibration Serum Level 3 (CAL 3) - multi-analyte mixture calibrator. Recalled by Randox Laboratories Ltd. Due to Randox can confirm a positive bias for Total...

Date: March 29, 2021
Company: Randox Laboratories Ltd.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Randox Laboratories Ltd. directly.

Affected Products

Calibration Serum Level 3 (CAL 3) - multi-analyte mixture calibrator. Catalogue Number: CAL2351

Quantity: 98 kits

Why Was This Recalled?

Randox can confirm a positive bias for Total Bilirubin and Direct Bilirubin of up to +15% across the assay ranges when using calibrator CAL2351, lot 1162UE. The positive bias has been attributed to bilirubin instability in the calibrator and is evident in both Quality Control (QC) and Patient samples.

Where Was This Sold?

This product was distributed to 15 states: CA, CT, IL, IN, MD, MA, MN, MS, NE, NJ, NY, OR, TX, VA, DC

Affected (15 states)Not affected

About Randox Laboratories Ltd.

Randox Laboratories Ltd. has 72 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report