Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Magnesium on RX Imola analyser IVD Recalled by Randox Laboratories, Limited Due to According to the firm, Carry over was observed...

Date: April 14, 2016
Company: Randox Laboratories, Limited
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Randox Laboratories, Limited directly.

Affected Products

Magnesium on RX Imola analyser IVD

Quantity: 15

Why Was This Recalled?

According to the firm, Carry over was observed when the amylase or pancreatic amylase test is run directly before or after the Magnesium assay. An update is required to the RX Imola analyser running order and an acid wash recommended when setting up the magnesium assay. A correction was made to the IFU for the Magnesium assay and all RX Imola customers in the USA were contacted with the updated instruction.