Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ABL90 FLEX PLUS Analyzer Recalled by Radiometer Medical ApS Due to The firm received reports of occurrences where the...

Date: February 3, 2020
Company: Radiometer Medical ApS
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Radiometer Medical ApS directly.

Affected Products

ABL90 FLEX PLUS Analyzer, Model Number 393-092 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.

Quantity: 3,968 units

Why Was This Recalled?

The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. This could result is patient mixup and/or lost of sample, resulting in delayed medical treatment.