Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Lactate Membrane units for E7077 Electrode Radiometer Medical ApS. The Recalled by Radiometer America Inc Due to RADIOMETER became aware that some membranes may have...

Date: October 8, 2012
Company: Radiometer America Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Radiometer America Inc directly.

Affected Products

Lactate Membrane units for E7077 Electrode Radiometer Medical ApS. The intended use is in vitro testing of samples of whole blood.

Quantity: 13,632 boxes/4 membranes

Why Was This Recalled?

RADIOMETER became aware that some membranes may have enzyme residue on the outer membrane. The enzyme residue may cause an initial negative bias on the reported Lactate result upon replacement of the Lactate membrane.

Where Was This Sold?

This product was distributed nationwide across all 50 states.