Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

ABL90 FLEX analyzers with software versions 3.3MR1 or below. The Recalled by Radiometer America Inc Due to The ABL90 FLEX can aspirate Cal 2 solution...

Date: May 25, 2018
Company: Radiometer America Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Radiometer America Inc directly.

Affected Products

ABL90 FLEX analyzers with software versions 3.3MR1 or below. The ABL90 FLEX Analyzer is a portable, automated Analyzer that measures pH, blood gases, electrolytes, glucose, lactate, bilirubin and oximetry in whole blood. It is intended for use in a laboratory environment, near patient or point-of-care setting.

Quantity: 12,102

Why Was This Recalled?

The ABL90 FLEX can aspirate Cal 2 solution instead of rinse solution, when the following take place: ¿ The analyzer is in ready mode ¿ Operator lifts the inlet and then closes it again, thereby initiating a rinse ¿ Immediately after lifts the inlet and quickly closes it again

Where Was This Sold?

US Distribution and Internationally to Canada.

About Radiometer America Inc

Radiometer America Inc has 25 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report