Algostim LLC Recalled by QiG Group LLC Due to The implantable pulse generator (IPG) model 2408 had...

Date: January 26, 2016
Company: QiG Group LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact QiG Group LLC directly.

Affected Products

Algostim LLC, Algovita Implanted Pulse Generator stimulator, Model 2408 ( 3 x 8 channel)

Quantity: 40

Why Was This Recalled?

The implantable pulse generator (IPG) model 2408 had suboptimal performance establishing a connection to the external devices unless they were brought into close proximity (as close as 5 cm). The following functions may not be available: Reprogramming the stimulator, Adjusting stimulation parameters (on/off, intensity modifications, stimulation program selection), and Recharging the stimulator.

Where Was This Sold?

Distributed in Germany and Luxembourg.