Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

EZ2 Connect MDx-IVD Designed to perform automated isolation and purification Recalled by Qiagen Sciences LLC Due to Insufficient welding leads to the heater cable becoming...

Date: March 4, 2024
Company: Qiagen Sciences LLC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Qiagen Sciences LLC directly.

Affected Products

EZ2 Connect MDx-IVD Designed to perform automated isolation and purification of nucleic acids in molecular diagnostic and/or molecular biology applications. Model Number: 9003230

Quantity: 168 units: 38 units US; 130 units OUS

Why Was This Recalled?

Insufficient welding leads to the heater cable becoming brittle and present with an error: ERROR 3481 HEATER UNABLE TO REACH TEMPERATURE; the instrument may stop, and the sample being processed will be lost

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Qiagen Sciences LLC

Qiagen Sciences LLC has 21 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report