Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Boppli: Bedside Device Kit Recalled by PYRAMES INC Due to Single-use battery-powered blood pressure monitor, for use on...

Date: February 24, 2025
Company: PYRAMES INC
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact PYRAMES INC directly.

Affected Products

Boppli: Bedside Device Kit, REF: 54-0002; Sensor Band, REF: 51-0007

Quantity: 35

Why Was This Recalled?

Single-use battery-powered blood pressure monitor, for use on neonates and infants, due to vibratory noise, may not adequately detect changes in blood pressure per specification. Containment actions: 1) Ventilator lines not to touch the isolette (e.g. mattress, side rails, pillow) or patient, 2) Move affected monitor to patient's foot.

Where Was This Sold?

This product was distributed to 2 states: CA, MA

Affected (2 states)Not affected