Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

MATResponder Tourniquet Recalled by Pyng Medical Corporation Due to Once tested, it is required that the tourniquet...

Date: August 31, 2015
Company: Pyng Medical Corporation
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Pyng Medical Corporation directly.

Affected Products

MATResponder Tourniquet; Catalog number MATR: UPC 78353 00008; MATCombat Tourniquet, Catalog number MATC; UPC 78353 00010; MATResponder Tourniquet and MATCombat Tourniquet are essentially the same product with different color schemes manual nonpneumatic tourniquet devices. The device provides for emergency blood flow occlusion to the limbs

Quantity: 6728

Why Was This Recalled?

Once tested, it is required that the tourniquet be returned to its Ready position then placed in its packaging. Each tourniquet was tested for functionality, but in some cases, the tourniquet was not returned to its Ready position prior to packaging.

Where Was This Sold?

Worldwide Distribution - US Distribution and to the countries of : Colombia, France, Canada, Italy, Germany, Finland, United Kingdom, Australia, The Netherlands, Japan, Hungary, Indonesia, Israel, Mexico, Poland and Spain.

About Pyng Medical Corporation

Pyng Medical Corporation has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report