Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Bayonet Forceps 8.5" (215.9mm) Serrated 1.0 mm Tips Electrosurgical/Monopolar Forceps/Reusable Recalled by PSC Industries Inc Due to Firm officials reported to CIN-DO that their Electrosurgical...

Date: June 1, 2012
Company: PSC Industries Inc
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact PSC Industries Inc directly.

Affected Products

Bayonet Forceps 8.5" (215.9mm) Serrated 1.0 mm Tips Electrosurgical/Monopolar Forceps/Reusable, Model #41604, 510 (k) #K885199. The surgical device is package in non-sterile small plastic tubes. Monopolar Reusable Forceps - are intended for use in cutting and coagulating tissue during surgical procedures.

Quantity: 5 units

Why Was This Recalled?

Firm officials reported to CIN-DO that their Electrosurgical Monopolar Reusable Bayonet Forceps are misbranded in that the label states that the tip size is: (0.5mm) when the tip size actually measures (1.0mm).

Where Was This Sold?

This product was distributed nationwide across all 50 states.