Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Posey Biothane Restraint Cuffs Recalled by J T Posey Company Due to The firm voluntarily recalled Biothane Restraint Cuffs, Catalog...

Date: October 31, 2012
Company: J T Posey Company
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact J T Posey Company directly.

Affected Products

Posey Biothane Restraint Cuffs, Catalog Number 2900. The device is intended to limit a patient's movement to the extent necessary for treatment or protection of the patient or others.

Quantity: 1673 units

Why Was This Recalled?

The firm voluntarily recalled Biothane Restraint Cuffs, Catalog No. 2900, that did not meet the required specifications for tensile strength. Failure of these rivets could allow the cuff to become disconnected from the connecting strap and allow patients to regain free use of the limb.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About J T Posey Company

J T Posey Company has 1 total recall tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report