PTS Panels Creatinine test strips. PTS Panels Creatinine Test Strips Recalled by Polymer Technology Systems, Inc. Due to During a routine product investigation, a creatinine sample...

Date: May 15, 2012
Company: Polymer Technology Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Polymer Technology Systems, Inc. directly.

Affected Products

PTS Panels Creatinine test strips. PTS Panels Creatinine Test Strips are intended to measure creatinine in whole blood. Creatinine measurements are used in the diagnosis and treatment of renal (kidney) diseases and in the monitoring of renal dialysis. This system is intended for the professional use for accurately and reproducibly measuring creatinine in human whole blood obtained from a finger stick sample

Quantity: 1650 vials

Why Was This Recalled?

During a routine product investigation, a creatinine sample with a value less than 1.0 mg/dl was tested on a version 2.60 CardioCek PA meter and the meter provided a result of 0.08mg/dl, which is not the reportable measuring range for the CardioChek PA meter for the creatinine assay.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Polymer Technology Systems, Inc.

Polymer Technology Systems, Inc. has 14 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report