Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

Henry Schein CardioChek Plus Professional Analyzer Recalled by Polymer Technology Systems, Inc. Due to One brand of batteries have a small dimensional...

Date: December 2, 2019
Company: Polymer Technology Systems, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Polymer Technology Systems, Inc. directly.

Affected Products

Henry Schein CardioChek Plus Professional Analyzer, New Analyzer, boxed, with carrying case, Reference Number 4060 HS, 570-0411

Quantity: 44,448 total

Why Was This Recalled?

One brand of batteries have a small dimensional difference in the negative button terminal, which, when the batteries are inserted incorrectly, can create the possibility for the battery terminals to make reverse contact. This reverse contact can cause a potential for overheating.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Polymer Technology Systems, Inc.

Polymer Technology Systems, Inc. has 14 total recalls tracked by RecallDetector.

Related Recalls