Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

The LIFEPAK 15 monitor/defibrillator is intended for use by trained Recalled by Physio-Control, Inc. Due to Customer complaint was received relating to a LifePak15...

Date: March 11, 2010
Company: Physio-Control, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Physio-Control, Inc. directly.

Affected Products

The LIFEPAK 15 monitor/defibrillator is intended for use by trained medical personnel in outdoor and indoor emergency care settings within the environmental conditions specified. Product Part Numbers V-15-2-000XXX

Quantity: 1399 units (866 in the US and 533 outside US)

Why Was This Recalled?

Customer complaint was received relating to a LifePak15 device that would power on then power down unexpectedly after a short period of use. When the LP15 powered down, power could be immediately restored by activation of the power button. The device was connected to a cellular modem from MultiTech systems for use with LIFE NET services. The corrective action was completed by Velocitor Solutions in December 2010.

Where Was This Sold?

worldwide

About Physio-Control, Inc.

Physio-Control, Inc. has 38 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report