Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

LIFEPAK 15 Monitor/Defibrillator with End-Tidal CO2 option. Intended for use Recalled by Physio-Control, Inc. Due to The End-Tidal CO2 (EtCO2) reading can intermittently show...

Date: January 16, 2017
Company: Physio-Control, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Physio-Control, Inc. directly.

Affected Products

LIFEPAK 15 Monitor/Defibrillator with End-Tidal CO2 option. Intended for use by trained medical personnel in out-of-doors and indoor emergency care settings within the environmental conditions specified in the Operating Instructions.

Quantity: 2584 units total (1501 units in the US; 1034 units international; and 49 units owned by Physio). 50 modules total

Why Was This Recalled?

The End-Tidal CO2 (EtCO2) reading can intermittently show a value of XXX after start-up or during device operation.

Where Was This Sold?