Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Trilogy EVO Ventilator Recalled by Philips Respironics, Inc. Due to Software defect in the Trilogy EVO and Trilogy...

Date: November 19, 2019
Company: Philips Respironics, Inc.
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Respironics, Inc. directly.

Affected Products

Trilogy EVO Ventilator, EU - Model Number: EU2110X15B Trilogy Evo is intended for pediatric through adult patients weighing at least 2.5 kg. The ventilator can measure, display, record, and alarm SpO2, FiO2, CO2, and Pulse Rate data when integrated with the appropriate accessories.

Quantity: 48

Why Was This Recalled?

Software defect in the Trilogy EVO and Trilogy EVO Universal Ventilator, versions SW 1.00.05, SW 1.01.09.00, SW 1.01.10.00, and SW 1.01.11.00, causing inoperative and loss of all power alarms. This defect causes an incorrect or unexpected result failing requirements TSRS1298 and PRD439.

Where Was This Sold?

This product was distributed to 1 state: DE

Affected (1 state)Not affected

About Philips Respironics, Inc.

Philips Respironics, Inc. has 85 total recalls tracked by RecallDetector.