Class I — Serious Health Hazard

Serious health hazard — there is a reasonable probability that use of or exposure to this product will cause serious adverse health consequences or death.

Philips Trilogy 200 Ventilator Recalled by Philips Respironics, Inc. Due to The replacement silicone sound abatement foam installed into...

Date: December 7, 2022
Company: Philips Respironics, Inc.
Status: Ongoing
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips Respironics, Inc. directly.

Affected Products

Philips Trilogy 200 Ventilator, Model Numbers 1040005, U1040005, 1040005B, U1040005B, CA1032800B, CA1032800, KO1032800, 1032800, R1032804TP, R1032804B, R1032804TPV, 1032804B

Quantity: 537 devices

Why Was This Recalled?

The replacement silicone sound abatement foam installed into some Trilogy 100 and Trilogy 200 devices may separate from the plastic backing to which it is attached. If this were to happen, the foam could potentially block air inlet, which could result in a reduction in delivered therapy volume or pressure and could cause the device to alarm.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Philips Respironics, Inc.

Philips Respironics, Inc. has 85 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report