Class II — Potential Health Hazard

Potential health hazard — use of or exposure to this product may cause temporary or medically reversible adverse health consequences.

URODiagnost Eleva Recalled by Philips North America, LLC Due to The locking plate may be missing from the...

Date: February 28, 2019
Company: Philips North America, LLC
Status: Terminated
Source: FDA (Device)

What You Should Do

Stop using this product immediately. Do not consume, use, or distribute it.

Return the product to the place of purchase for a full refund. If you have questions, contact Philips North America, LLC directly.

Affected Products

URODiagnost Eleva, Product codes 708026, 708027, 708028 Product Usage: The Philips OmniDiagnost is a multi-functional, universal X-ray system with remote-controlled facilities. It performs a wide range of routine examinations and interventional procedures include: Gastro-intestinal examination Urography Bronchography Arthrography Hysterography Myelography Lymphography Skeletal radiography Tomography Venography/phlebography Special procedures that require subtracted and non-subtracted digital imaginginclude: Angiography Peripheral angiography (bolus chasing) Vascular and non-vascular interventional procedures.

Quantity: 1969

Why Was This Recalled?

The locking plate may be missing from the system, which may result in unexpected stand/table tilt movement that cannot be stopped by the user.

Where Was This Sold?

This product was distributed nationwide across all 50 states.

Affected (51 states)Not affected

About Philips North America, LLC

Philips North America, LLC has 88 total recalls tracked by RecallDetector.

Related Recalls

Data sourced from the FDA (Device). Last updated March 26, 2026. View original report